Class I Recall Issued for Automated Impella Controllers by Abiomed

The US Food and Drug Administration has classified the removal of certain Automated Impella Controllers by Abiomed as a Class I recall, which is the most serious type of device recall. A retrospective review of servicing records identified that these controllers require specific hardware, firmware, or software updates to address quality issues that could carry safety implications. Identified issues include potential failure to recognize the purge cassette during setup or changes, unreliable buttons that could affect menu selections and console control, and start-up instability.

Without the necessary updates, patients could experience a delay in therapy, as well as inadequate or lost hemodynamic support. Abiomed has noted that current hospital inventory can continue to be used while awaiting service to maintain continuity of care. The manufacturer's service team will coordinate with facilities to schedule the removal, updating, and return of affected devices.

Key points

  • Abiomed is removing certain Automated Impella Controllers for required hardware, firmware, and software updates.
  • Identified issues include purge cassette recognition failure, sticking buttons, and start-up instability.
  • Potential risks include delayed therapy and loss of hemodynamic support.
  • Existing hospital inventory may continue to be used while awaiting service to ensure continuity of care.

What it means for you

Clinicians and healthcare staff utilizing these cardiac support devices should be aware of the potential risks of delayed therapy or console control issues. Facilities should coordinate with the manufacturer for necessary device updates while utilizing current inventory for patient care continuity. Clinicians in Pakistan should consult DRAP or local suppliers regarding the availability and safety status of these specific controllers in local clinical settings.