Early Alert Issued for Ventilator Issue with React Health VOCSN Devices

The US Food and Drug Administration (FDA) has issued an early alert regarding a potentially high-risk device issue affecting all VOCSN ventilator devices with serial numbers 112786 and higher manufactured by Ventec Life Systems, doing business as React Health. The internal flow transducer cable connector in these devices may become unseated and lose communication with the internal flow transducer board, leading to device reboots and recurring System Fault 10 alarms.

React Health has issued a letter to affected customers recommending updated use instructions and inspection procedures. If a device exhibits this issue, users should utilize back-up ventilation and contact the company for servicing. Patients are advised not to discontinue therapy unless instructed by their treating clinician and an alternative means of ventilation is available.

Key points

  • Internal flow transducer cable connectors may unseat, causing device reboots and System Fault 10 alarms.
  • The issue affects all VOCSN ventilator device configurations with serial numbers 112786 and higher.
  • If the device exhibits the issue, switch to back-up ventilation and contact React Health for servicing.
  • Ensure alternative ventilation equipment is functional, regularly checked, and ready for immediate use.

What it means for you

Clinicians and healthcare providers should promptly notify affected patients and caregivers regarding this device issue and verify that reliable backup ventilation is available. As this is an international FDA alert, medical professionals in Pakistan should check with local suppliers or the Drug Regulatory Authority of Pakistan (DRAP) to determine if these specific React Health VOCSN models are registered, imported, or in use locally.