FDA Alerts Clinicians to Removal of Portex Endotracheal Tubes Due to Material Softness

The US Food and Drug Administration has issued an early alert regarding ICU Medical's removal of specific Portex Endotracheal Tubes from clinical and distribution settings. The affected devices comprise selected uncuffed 2.5 mm, 3.0 mm, and 3.5 mm Murphy eye oral or nasal tubes, alongside 3.0 mm and 3.5 mm Polar clear south oral uncuffed preformed tubes.

The action follows the identification of a manufacturing issue involving resin with variable plasticizer percentages, leaving the tubes softer than intended. In neonatal and early infant populations, this softness can trigger tube kinking, elevated resistance during ventilation, and intubation delays. If undetected, the defect may cause inadequate ventilation, prompt unscheduled tube replacements, and contribute to serious adverse events, including one reported death.

ICU Medical advises healthcare institutions to immediately inspect all inventory, discontinue using affected devices, and destroy or quarantine them to prevent clinical use. Facilities are also advised to share the alert with any secondary recipients and report device-related adverse events.

Key points

  • ICU Medical has initiated the removal of specific uncuffed 2.5 mm, 3.0 mm, and 3.5 mm Portex endotracheal tubes.
  • A resin manufacturing defect causes the tubes to be abnormally soft, risking tube kinking, ventilation resistance, and intubation failure.
  • The issue affects neonatal and infant patients, with one death reported in connection with the defect.
  • Healthcare providers must check inventory immediately, cease use, and destroy or quarantine all affected products.

What it means for you

Clinicians, anesthesiologists, and pharmacy teams in Pakistan should inspect pediatric and neonatal airway inventories to confirm whether any affected Portex endotracheal tubes are in stock. Facilities using imported airway devices should consult DRAP and their authorized distributors to determine local product status and quarantine any matching lots. Suspected device malfunctions or adverse events during intubation should be isolated and reported in accordance with institutional and regulatory procedures.

Read the original announcement

Early Alert: ICU Medical Removes Portex Endotracheal Tubes