FDA Alerts Healthcare Providers to Valve Malfunction in B. Braun Caresite Micro Access Devices

The US FDA has released a safety alert concerning B. Braun Medical Inc.'s Caresite Micro Luer Access Device, component number S5404000. The manufacturer issued an urgent correction after discovering that the internal piston can stick in a depressed position upon disconnecting a male luer connector. An absence of required silicone lubricant prevents the valve from closing properly, which can lead to backflow, fluid leakage, and air ingress. A valve that remains open can result in severe clinical complications, including blood loss, therapy interruption, under-infusion of critical medications, and air embolism. Vulnerable groups such as neonates, pediatric patients, and critically ill individuals are at heightened risk of harm from these events, though no serious injuries or deaths have been reported to date. Healthcare facilities are instructed to train staff to inspect every valve immediately after disconnecting a connector. If backflow, leakage, or an open valve is observed, clinicians should promptly intervene to stop fluid loss, assess the patient for blood loss or therapy disruption, and replace the affected device using aseptic technique.

Key points

  • B. Braun initiated an urgent medical device correction for Caresite Micro Luer Access Devices with component number S5404000.
  • Missing silicone lubricant may cause the internal piston to stick open following male luer disconnection.
  • Open valves create risks of blood or medication leakage, infusion delays, and potential air embolism.
  • Clinical staff must visually inspect the access device after each disconnection and replace it immediately if leakage or an open state is seen.
  • No serious injuries or deaths have been reported to date in connection with the device issue.

What it means for you

Clinicians and hospital pharmacists in Pakistan should check their facilities' inventories for the affected B. Braun access devices and consult DRAP for local regulatory status and safety notifications. Healthcare personnel using needle-free connectors should maintain strict vigilance upon disconnection, immediately inspecting for valve closure and replacing faulty devices per standard aseptic protocols.

Read the original announcement

Early Alert: B. Braun Medical Inc. Updates Use Instructions for Caresite Micro Luer Access Device