FDA Alerts on Recall of Tandem Mobi Insulin Pumps Due to False Occlusion Alarms

The US Food and Drug Administration has highlighted a Class I recall by Tandem Diabetes for Tandem Mobi Insulin Pumps with Interoperable Technology. The recall requires removing and replacing certain pumps based on their serial number, while providing updated operating instructions for remaining devices. The action was initiated because affected pumps may generate false-positive occlusion alarms when no physical blockage exists, resulting in an improper suspension of insulin delivery until acknowledged by the user.

Repeated or unnoticed suspensions of insulin delivery can lead to elevated blood glucose levels, potentially causing severe hyperglycemia that requires medical intervention or hospitalization. Tandem Diabetes has received reports of 1,248 serious injuries and no deaths connected to this issue.

Customers with pumps flagged for replacement must immediately submit a request for a replacement unit from the manufacturer. All users of Tandem Mobi pumps are advised to respond promptly to occlusion alarms, inspect infusion sites and tubing, enable smartphone push notifications via the mobile application, monitor blood glucose closely, and ensure a backup insulin delivery plan is readily available.

Key points

  • Tandem Diabetes is recalling and replacing select Tandem Mobi insulin pumps due to false-positive occlusion alarms that halt insulin delivery.
  • The defect has resulted in 1,248 reported serious injuries and no deaths related to hyperglycemia from suspended insulin administration.
  • Affected devices are categorized by serial number into those requiring immediate replacement and those requiring updated use instructions.
  • Users of all Tandem Mobi pumps must enable mobile app alerts, troubleshoot alarms promptly, and keep backup insulin delivery methods on hand.

What it means for you

Healthcare providers in Pakistan managing patients using Tandem Mobi insulin pumps should advise them to verify their pump serial numbers and seek replacements if affected. Clinicians should reinforce the importance of responding immediately to occlusion alarms, monitoring blood glucose, and maintaining accessible backup insulin delivery supplies. Consult the Drug Regulatory Authority of Pakistan for information regarding local registration and authorized device distribution.

Read the original announcement

Insulin Pump Recall: Tandem Diabetes Removes Tandem Mobi Insulin Pumps

Summarised automatically from the official announcement. For educational purposes only — check the original source before making clinical decisions.