Intersurgical Recalls 22mm Smoothbore Breathing Tubes and Systems

Intersurgical Inc. has issued a recall for specific lots of 22mm Smoothbore breathing tubes and systems after investigations revealed that soft connectors may detach during use due to insufficient solvent application during manufacturing.

The affected devices carry air and oxygen to patients and allow exhaled air to flow out. If a connector detaches, it can lead to a loss of ventilation, potentially causing patient desaturation, hypoxia, organ failure, brain damage, or death.

The manufacturer advises customers to immediately stop using and quarantine any affected stock, return the products to Intersurgical, and share the recall notice with all facility users and customers who received the products.

Key points

  • Recalled devices include various lots of Intersurgical 22mm Smoothbore breathing tubes and systems.
  • Soft connectors may disconnect from the tube due to insufficient Tetrahydrofuran solvent used during manufacturing.
  • Detachment can cause loss of ventilation, leading to hypoxia, brain damage, or death.
  • Users are advised to immediately stop using, quarantine, and return affected lots.

What it means for you

Clinicians and clinical staff should immediately check their inventory for the recalled Intersurgical 22mm Smoothbore breathing tube and system lots, quarantine any affected devices, and cease their use to prevent ventilation failure. Healthcare facilities in Pakistan should verify local stock availability and consult DRAP for any further context regarding imported medical devices.

Read the original announcement

Breathing Systems Recall: Intersurgical Inc. Removes 22mm Smoothbore Breathing Tubes and Systems