Recall Correction Issued for Luna G3 BPAP System Over Inaccurate Therapy Pressure Data

The US Food and Drug Administration has published a voluntary medical device correction notice from React Health concerning specific Luna G3 Bilevel Positive Airway Pressure (BPAP) systems, model LG3800-30VT. The recall involves 7,533 devices distributed in the United States and Puerto Rico due to a data storage and transmission error that occurs when the Targeted Tidal Volume (Targeted VT) feature is activated.

When Targeted VT is enabled, the device fails to record or wirelessly transmit actual delivered therapy pressure statistics to its internal SD card or the React Health Connect cloud platform. While the device continues to deliver ventilation within clinician-set tidal volume and pressure limits, the inaccurate historical reports may lead clinicians to adjust therapy settings inappropriately or experience delays in optimizing care. React Health has received fifteen complaints and two serious adverse reports associated with the issue.

React Health has released firmware version G3-2.00.80 to correct the reporting defect. Clinicians and equipment providers are advised to discontinue using Targeted VT on affected units, quarantine unassigned inventory, and carefully re-evaluate historical patient data. Patients are cautioned not to alter device settings or stop prescribed therapy without clinical guidance.

Key points

  • React Health has issued a voluntary recall correction for Luna G3 BPAP Model LG3800-30VT devices due to data reporting errors.
  • When Targeted VT is enabled, the device does not properly store or transmit accurate delivered pressure statistics.
  • Fifteen complaints and two serious adverse reports linked to the reporting issue have been documented.
  • Clinicians should immediately disable Targeted VT on affected machines until firmware version G3-2.00.80 is installed.
  • Patients should continue their prescribed respiratory therapy and contact their provider rather than stopping or altering settings independently.

What it means for you

Clinicians and biomedical departments in Pakistan using imported Luna G3 BPAP systems should verify whether model LG3800-30VT is deployed with Targeted VT enabled. Practitioners should avoid relying on historical pressure logs from affected machines to make clinical decisions until the manufacturer's firmware update is applied, and consult DRAP for any corresponding local alerts.